Calibration
Selecting a Calibration Mode
Note:
• All calibration procedures must be performed by trained person-
nel. Incorrectly set parameters may go unnoticed, but change the
measuring properties.
Calibration can be performed by:
• Determining the cell constant with a known calibration solution
• Input of cell constant (e.g. for ultrapure-water sensors)
• Entering an installation factor*
)
• Sampling (product calibration)
• Temperature probe adjustment
Calibration adapts the device to the individual sensor characteristics.
Access to calibration can be protected with a passcode
(SERVICE menu).
First, you open the calibration menu and select the calibration mode:
Calibration with calibration solution
Calibration by entry of cell constant
Calibration by entry of an installation factor*
)
Product calibration (calibration with sampling)
Temperature probe adjustment
*) with Memosens sensors only