55
Switching Parameter Sets A/B
Note:
With Cond-Cond, the two parameter sets are replaced by the two sensors A and B.
External switchover of parameter sets A/B
You can switch between parameter sets A and B by applying a signal to the
CONTROL input (setting:
).
Parameter set A active
0...2 V AC/DC
Parameter set B active 10...30 V AC/DC
Power supply (PWR out)
12...24 V AC/DC
Stratos Evo A402
13
12
CONTROL
input
Parameter Set A/B: Configurable Menu Groups
The device provides 2 parameter sets “A“ and “B“. By switching between the parameter sets you can adapt
the device to different measurement situations, for example. Parameter set “B“ only permits setting of
process-related parameters.
Menu group Parameter set A
Parameter set B
Sensor selection
---
Current output 1
Current output 1
Current output 2
Current output 2
Compensation
Compensation
Control input
---
Alarm mode
Alarm mode
Relay outputs
Relay outputs
Cleaning (not for Cond-Cond
devices)
---
Parameter set selection
(not for Cond-Cond devices)
---
Setting the clock
---
TAG of measuring point
GROUP of measuring points