67
Cond Configuration
Menu item
Action
Choices
CIP / SIP
Cleaning cycles on/off
Select ON or OFF using
keys.
Activates/deactivates logging in
extended logbook
Press enter to confirm.
Sterilization cycles on/off
Select ON or OFF using
keys.
Activates/deactivates logging in
extended logbook
Press enter to confirm.
The cleaning and sterilization cycles are logged to measure the load on the sensor.
Suitable for biochemical applications (process temperature approx.
0 ... +50 °C / +32 ... +122 °F, CIP temperature > +55 °C / +131 °F,
SIP temperature > +115 °C / +239 °F).
Note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure
that the cycle is complete.
With Memosens sensors, an entry is also made in the sensor.
5