160
Error Messages
Error
Info text
(is displayed in case of fault when
the Info key is pressed)
Problem
Possible causes
ERR 95 SYSTEM ERROR
System error
Restart required.
If error still persists, send in the
device for repair.
ERR 96 WRONG MODULE
Module does not correspond
to measuring function
Correct the setting in the
SERVICE / DEVICE TYPE menu.
Afterwards, configure and
calibrate the device.
ERR 97 NO MODULE INSTALLED
No module installed
ERR 98 CONFIGURATION ERROR
Error in configuration or
calibration data
Configuration or calibration data
defective; completely reconfig-
ure and recalibrate the device.
ERR 99 DEVICE FAILURE
Error in factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect.
The device must be repaired and
recalibrated at the factory.