159
Error Messages
Error
Info text
(is displayed in case of fault when the
Info key is pressed)
Problem
Possible causes
ERR 59 INVALID CALCULATION
Invalid calculations
ERR 60 OUTPUT LOAD
Load error
ERR 61 OUTPUT 1 TOO LOW
Output current 1
< 0 (3.8) mA
ERR 62 OUTPUT 1 TOO HIGH
Output current 1
> 20.5 mA
ERR 63 OUTPUT 2 TOO LOW
Output current 2
< 0 (3.8) mA
ERR 64 OUTPUT 2 TOO HIGH
Output current 2
> 20.5 mA
ERR 72 FLOW TOO LOW
Flow too low
ERR 73 FLOW TOO HIGH
Flow too high
ERR 74 CATION EXCHANGER INVALID
CALCULATION
Flow too low or no flow:
Flow ≤ 4.00 l/h;
calculated pH value:
< 7.5 or > 10.5;
conductivity values: B ≥ 3 x A
Содержание Stratos Evo A402