155
Menu item
Remark
Assigning passcodes:
In the “SERVICE - CODES“ menu you can assign pass-
codes to DIAG, HOLD, CAL, CONF and SERVICE modes
(Service preset to 5555).
When you have lost the Service passcode, you
have to request an “Ambulance TAN“ from the man-
ufacturer specifying the serial number and hardware
version of your device.
To enter the “Ambulance TAN“, call the Service func-
tion and enter passcode 7321. After correct input of
the ambulance TAN the device signals “PASS“ for 4 sec
and resets the Service passcode to 5555.
Reset to factory settings:
In the “SERVICE - DEFAULT“ menu you can reset the
device to factory settings.
NOTICE!
After a reset to factory setting the device must
be reconfigured completely, including the sensor
parameters!
Option request:
Communicate the serial number and hardware/soft-
ware version of your device to the manufacturer.
These data can be viewed in the Diagnostics/Version
menu.
The “transaction number“ (TAN) you will then receive
is only valid for the device with the corresponding
serial number.
Releasing an option:
Options come with a “transaction number“ (TAN).
To release the option, enter this TAN and confirm by
pressing enter.
Device type:
Changing the measuring function, e.g., after having
replaced a Memosens sensor.
Not possible with a measuring module installed.
Service