153
Service
Menu item
Remark
Displaying currently measured values (sensor
monitor) with HOLD mode activated:
Select MONITOR using
, press enter to confirm.
Select variable in the bottom text line using
.
The selected parameter is shown in the upper dis-
play line.
As the device is in HOLD mode, you can perform
validations using simulators without influencing the
signal outputs.
Hold meas depressed for longer than 2 sec to
return to Service menu.
Press meas once more to return to measurement.
NEW EXCHANGER
When you have replaced the ion exchanger, select
“YES” to reset the consumption calculation.
POWER OUT
POWER OUT, adjusting the output voltage
Here, you can select an output voltage of
3.1/12/15/24 V.