5
Intended Use
SensoGate® WA 131 H
Intended Use
The SensoGate® WA 131 H hygienic sensor lock-gate is used for installing a sensor for measure-
ments in liquids. The sensor can be cleaned, calibrated or replaced under process conditions
(pressure and temperature).
The operator can exchange process adaptations (Ingold socket, dairy pipe, Varivent, BioControl,
Clamp) or convert the fitting for the use with gel sensors or liquid-electrolyte sensors.
The sensor lock-gate is suitable for sensors with an outer diameter of 12 mm:
• with gel electrolyte, length 225 mm, sensor head with PG 13.5
• with liquid electrolyte, length 250 mm
The SensoGate® WA 131 H sensor lock-gate allows:
• calibrating or adjusting the measuring system and cleaning the sensor in the running process
• replacing the sensor in the running process
• variable process adaptation by the operator at any time
Take account of the influences of humidity, ambient temperature, chemicals and corrosion.
Safe use
If you are not sure whether the sensor lock-gate can be safely used for your intended
application, please contact the manufacturer.
To ensure safe use of the equipment, you must follow the instructions given in this manual and
observe the specified temperature and pressure ranges.
The SensoGate® WA 131 H sensor lock-gate has been developed and manufactured in compli-
ance with the applicable European guidelines and standards. Compliance with the European
Harmonized Standards for use in hazardous locations is confirmed by the EC-Type-Examination
Certificate. Compliance with the European guidelines and standards is confirmed by the EC
Declaration of Conformity.
For hazardous-area applications, the sensors used must ensure proper separation of the ATEX
zones.
There is no particular direct hazard caused by the operation of the device in the specified
environment.
Hygienic Design
With regard to hygienic design and sterilizability, the retractable fitting complies with the
recommendations of EHEDG. This was established and verified in the TNO report V7942 dated
February 25, 2008. If required, the TNO report can be viewed or obtained from the manufacturer.
Содержание SensoGate WA 131 H
Страница 1: ...SensoGate WA 131 H User Manual Sensor Lock Gate ...
Страница 46: ...46 Notes SensoGate WA 131 H ...
Страница 47: ...47 Notes SensoGate WA 131 H ...