![Knick SensoGate WA 131 H Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/knick/sensogate-wa-131-h/sensogate-wa-131-h_user-manual_1980170024.webp)
24
Installing the Immersion Tube
1) Push the O-ring
(D)
(sensor gasket) onto the
sensor
(E)
as shown.
Note:
Make sure that there is no further
O-ring in the immersion tube
(T)
(installed
by mistake). (For O-ring dimensions, see
page 43 “Sealing Kits for Maintenance and
Servicing”.)
2) Loosen the two screws
(Z)
by approx.
4 turns (do not detach them) if you have not
done that when removing the immersion
tube.
3) Push the immersion tube
(T)
in direction
of the arrow and insert it in the bayonet
coupling
(Y)
.
4) Press the tube firmly in place and turn it
clockwise until the stop (approx. 60°).
5) Fasten the two screws
(Z)
using a screw-
driver (TX25).
T
D
E
Z
Y
Z
Note:
The bayonet coupling is locked by the
form-fit screw heads. The immersion tube,
however, remains movable to compensate
for tolerances.
Screwdriver no. 6
included in accessory:
ZU 0680, ZU 0740, ZU 0754