17
Installing the Sensor
1) The sensor must only be installed in
SERVICE position
.
2) Use appropriate sensors
(A)
only:
Diameter: 12 mm Length: 225 mm
Observe pressure resistance of the sensor.
3) Check whether slide washer
(C)
or O-ring
(D)
(or HD gasket) on the sensor are missing or
damaged.
4) Screw in the sensor head
(B)
(19 mm A/F, PG 13.5) with a max. torque of 3 Nm
(recommended tool: 19 mm, e.g., Knick ZU0647
wrench).
5) Connect cable socket with cable
(G)
. Hold the
cable in a loop and fix it using clamp
(E)
.
CAUTION!
The cable loop must be long enough
so that the cable does not impede the stroke
movement of the fitting.
6) Connect equipotential bonding cable to terminal
(F)
(if required).
7) Mount protective cap (ZU 0759) if required.
Refer to manual for protective cap for mounting
instructions.
A
B
C
D
E
F
G
Installing a Gel-Electrolyte Sensor
Short immersion depth
ZU 0647