28
Connecting the HPLC Pump K–1001 to Other Instruments
3. Adjust the distances between this brackets be to the instrument below and
tighten the screws carefully.
The pump must be turned off and separated from the network before the
solvent organizer is connected or disconnected to the pump! Inserting or
detaching the connector with the pump switched on may seriously
damage the pump!
SOP 21 Connecting the Solvent Organizer K–1500
This SOP refers to the WellChrom HPLC Pump K-1001 and the Solvent
Organizer K–1500. Use this SOP to connect the Solvent Organizer to the HPLC
Pump K–1001.
1. Place the Solvent Organizer near your pump.
2. Connect the enclosed connection cable at the rear panel of the pump into
the „Plug for interconnection of Solvent Organizer K-1500“, Pos. {5.10},
see Fig. 4 on page 10.
3. Remove both stainless steel pans from the Solvent Organizer, Pos. {6.1}
Fig. 21.
4. Locate the logic board behind the Solvent Organizers front panel.
5. Insert cable through round hole in the rear panel of the Solvent Organizer
and connect cable with appropriate plug at the logic board.
6. Insert both pans for solvent bottles.
7. Connect solvent tubings (1/8" outer diameter) with nuts and fittings.
Unused channels of the Solvent Organizer may be left open.
During operation the red „LEDs A to D“, Pos. {6.2} indicate the currently open
valve.
Electrical Connections
Using the EVENT and ANALOG/ERROR/START Sockets
Electrical signal lines from other instruments can be connected to
the Analog/error/start or
Events socket or the
RS232 serial interface
on the rear side of the HPLC Pump K–1001, see positions {5.2},{5.4}, and {5.5}
in Fig. 4 on page 10.
Events can be controlled in the main menu or in programs. You can use SOP 9
„Setting Events“ on page 19.
Do not touch the electrical contacts of the socket lines. Electrostatic
discharges can lead to damage of the electronics of the HPLC Pump K–
1001.
Configuring Plug Strips
For external control of the HPLC Pump K–1001 you have to configur a special
plug strip according to the local conditions. You can use the plug strips with 4 or
10 connectors, enclosed in the accessories, for easy installation of electrical
connections.