24
Programming the HPLC Pump K–1001
Example for creating a program
Generation of program no. 7 with the following functions: Constant
fl
ow rate of
100.0 ml/min. Linear gradient changing composition from 90 % A and 10 % B at
start time to 10 % A and 90 % B at time 20 minutes.
This program is based upon the configuration in section „Operating the HPLC
Pump K–1001“ on page 17. This setup must be performed before program
definition. Remember, underscored characters indicate that the corresponding
parameter is continuousy calculated.
1. Select the operation screen
2. Move cursor to the field Edit Prog, press 7 to generate program no. 7.
3. Press
À
to enter the Edit mode for the new program.
The cursor is positioned at Time 000.0.
4. Press
¿
to set the flow rate: Press 1, 0, 0 and then
À
to enter the value
100.0 ml/min. The display should look like part A see Fig. 16 on page 22,
cursor blinking in the TIME field.
5. Press
¾
to enter field % A , then 9 and 0 and
½
to confirm the input. The
amount for % B is automatically calculated and indicated as 10 %. The
display should now look like part B in the figure.
6. Press
now
À
again to create the next program line by defining a new
time step. The display now should look like part C in the figure, the cursor
blinking on the first asterisk.
7. Now press 2 and 0 and then
¾
to define time step 20.0 min as well as to
enter field % A.
8. Press 1 and 0 for 10 % A and then
½
to confirm the input. The amount for
% B is automatically calculated and indicated as 90 %. Part D in the
figure shows the current situation.
After loading, the program can now be executed.
Using Program Links
Program links are combinations of existing programs, which can be created and
edited like any programs.
Links use reserved program numbers
21 … 29
. These numbers are
automatically converted like
21 = L1. 22 = L2
etc. The HPLC Pump K–1001 can
store up to 9 different links.
Links use line numbers from
No01
to
No99
, each line referencing a linked
program. The parameter
R
demands the number of runs to be performed with
that program. A wait status can be programmed. The maximum number of
programs referenced in all links is limited to 100.
Any program can be referenced multiple times in a link. A program link can not
reference other links.
SOP 17 Creating Program Links
This SOP applies to the HPLC Pump K–1001 Firmware Revision 6.0X. Use this
SOP for creating program links in your HPLC Pump K–1001.
You can create links between existing programs as follows:
1. Enter desired number for link in field EDIT PROG and press
¾
for
confirmation.
2. Press any numeric key to create a line number. Lines are numbered
automatically and can be interchanged by renumbering.