12
Product information
4000 MiD® Instructions, V6695
dialog in the software. The LED will not be reset until the application is
restarted, which is also required to confirm the fault has been remedied.
Orange: Indicates that the MiD is acquiring data (mass scanning). The
LED will not be lit unless the MiD is scanning, for example it will not be lit
when a start or stop time is in effect.
Green: The hardware interlock is engaged when the LED is lit and vice
versa, see „5.8.4 Operating the front cover safety interlock“ on page 26
for more details.
Yellow: Indicates that the MiD is in operate mode, when unlit the MiD is in
standby.
3.2.3 The rear panel
Fig. 3 shows the connections and ports on the rear panel of the MiD.
These ports include terminals and connectors for the gas supply, vacuum
exhaust system and computer peripherals.
Refer to the labels listed in the table below when connecting the MiD and
any associated instruments.
Warning
The 4000 MiD provides double/reinforced insulation
to mains circuits. Any external equipment that is connected to the
4000 MiD through the rear panel must itself also provide double/
reinforced insulation to the mains supply.
Avertissement
Le 4000 MiD offre une double isolation renfor-
cée pour l’alimentation secteur. Tout équipement externe connec-
té aux 4000 MiD par le panneau arrière doit lui aussi fournir une
double isolation renforcée à l’alimentation secteur.
Connection
Description
A
Nebuliser Gas Port
B
Aux I/O Port
C
VGA Monitor Port
D
RS232 Port
E
4x USB Ports
F
2x Ethernet Ports
G
Pump exhaust
H
Main power
I
PS/2 Ports
J
eSATA Port
K
HDMI Port