98
PFX2512
SL01-PFX 8Slot Unit
The 8Slot Unit SL01-PFX is connected to the PFX2512 Charge/Discharge System Controller to
expand the voltage measurement points. For this connection, an EX01-PFX connection board is
installed into the PFX2512.
It enables highly accurate evaluation of cell voltage disparity measurements, which is
indispensable for evaluation testing of large capacity battery modules.
If Voltmeter Units OP03-PFX are installed in all SL01-PFX slots, voltage measurement points can be
expanded to 64 points.
Further, by installing Volt / Thermometer Units OP02-PFX in the PFX2512, you can increase the
number of measurement points to 72.
The DUT (battery)
OP02-PFX
OP03-PFX
SL01-PFX
Up to 2
(when an EX01-PFX is installed)
Up to 8
Cell voltage 8 points
Thermocouple 8 points
200 A
Charge/discharge
LAN
LAN
LAN
Cell voltage 64 points
EX01-PFX
PFX2512
Switching hub
Application software
BPChecker3000
8Slot Unit (Option) / Voltmeter Unit (Option)