PFX2512
Troubleshooting
The correct temperature is not displayed.
Tests cannot be performed properly.
Symptom
Items to check and possible remedy
The display error is large.
Are you using the specified thermistor? If a capacitor or other
component is connected to the thermistor, display error will become
large. Use the specified thermistor.
Check how the thermistor is connected to the DUT. Even a minute gap
will greatly affect measurement results.
See
Symptom
Items to check and possible remedy
Even when channels are
searched for, none is found.
If you are using a new PC or Windows, Windows Update may not be
installed.
Perform Windows Update to install the latest update.
If proper operation is verified, you do not need to perform additional
updates.
—
Tests do not start.
Is the model ID set correctly?
Is the ALARM/WARNING LED lit orange?
The vibration sensor may be activated.
Check whether the product is installed on a horizontal stable surface.
Is the cable connected to the DC power supply and Electronic load
properly connected?
CC/CV operation is strange.
Is a DUT other than a battery connected?
If a DUT other than a battery is connected, the CC/CV control may not
work properly.
If there is no load (not connected) or if a battery with an abnormally
high internal impedance is connected, CV mode may not work
properly.
Is voltage sensing connected?
The system will not work properly if it is not connected. Be sure to
connect the voltage sensing cable.
The DC power supply circuit
breaker trips.
The DC power supply’s protection function (OVP or OCP) may have
been activated. Check the DC power supply’s protection function
settings.
This behavior may also occur when there is no load or when current is
not flowing through the DUT (battery) (protection circuit activation
and the like).
Even though option boards are
installed, the number of
measurement points has not
been expanded.
Are option boards inserted properly?
Have you customized the model ID?
The first time that you use the product after purchase, use
BPChecker3000 to customize the model ID.
Io Config
Did you set the number of OP02-PFX or OP03-PFX boards?
Set the number of boards using BPChecker3000.
Test
Condition Editor
See
?
Help
?
Help
Содержание PFX2500 Series
Страница 12: ...12 PFX2512 This page is intentionally blank ...
Страница 27: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2512 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Страница 96: ...96 PFX2512 This page is intentionally blank ...
Страница 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Страница 143: ......