22
PFX2512
Combination for Performing Charge/Discharge Tests
0
10
20
30
0 5 10 15 20 25 30 35 40 45 50
40
50
60
70
Charge operation range of the DC power supply [PWR-01L]
Current (A)
Voltage (V)
DUT cable: At nominal cross-sectional area 14 mm
2
5 m
PWR401L
PWR801L
PWR1201L
PFX2512 Operating range
0
10
20
30
0 5 10 15 20 25 30 35 40 45 50
40
50
60
70
Charge operation range of the DC power supply [PWR-01ML]
Current (A)
Voltage (V)
DUT cable: At nominal cross-sectional area 14 mm
2
5 m
PWR401ML
PWR801ML
PWR1201ML
PFX2512 Operating range