44
PFX2512
Connecting the DUT (Battery)
Connecting the PFX2512 to the DUT (Battery)
1
Check that the PFX2512 is ready to execute tests (power-on state).
You can connect or disconnect DUT cables and sensing connectors even when the
power switch is on. But, make sure that a charge/discharge test is not in progress.
2
Connect DUT cables to the positive and negative DUT terminals of the input/
output terminal board of the PFX2512.
Connect the polarities according to the colors that you decided to use when creating
the cables.
Run the cables out as horizontal as possible, and fix them in place with screws.
3
Connect the other end of the DUT cables to the DUT (battery).
Connecting sensing cables
1
Insert the assembled sensing connector into the PFX2512 SENSING connector.
2
Connect or attach a voltage sensing cable and thermistor to the DUT
(battery).
DUT cable (+)
DUT cable (–)
Screw (M6x14)
Tightening torque: 3.0 N
・
m
To DUT (battery)
– terminal
To DUT (battery)
+ terminal
Содержание PFX2500 Series