PFX2512
73
PFX2512 Functional Specifications
5
Specific
a
tions
Discharge function
Static
Constant current
discharge
CC
Setting
Constant current value (Current)
Cutoff
condition
Specified time after discharge start (Discharge Time)
Battery voltage (Cutoff Voltage)
Capacity calculation voltage (Capacity Voltage)
1
1.
When the battery voltage goes below the capacity calculation voltage, discharge capacitance integration is
stopped.
Battery temperature (Max Temp)
Discharge capacity (Discharge Ah)
Rest
Specified time after discharge end (Rest Time)
Rest completion condition of the battery temperature (Rest Temp)
Constant voltage-
constant current
discharge
CC - CV
Setting
Constant current value (Current)
Constant voltage value (CV Voltage)
Constant cell voltage value (Cell CV Voltage)
2
2.
Can be set only when the optional OP02-PFX Volt/Thermometer Unit or OP03-PFX Voltmeter Unit is
installed.
Cutoff
condition
Specified time after discharge start (Discharge Time)
Discharge capacity (Discharge Ah)
Battery temperature (Max Temp)
Specified time after constant voltage operation (CV Time)
Specified current after constant voltage operation (It Current)
Specified time after current falls below It Current (It Time)
Rest
Specified time after discharge end (Rest Time)
Rest completion condition of the battery temperature (Rest Temp)
Constant power
discharge
CP
Setting
Constant power value (Wattage)
Limit current (Limit Current)
Cutoff
condition
Specified time after discharge start (Discharge Time)
Battery voltage (Cutoff Voltage)
Capacity calculation voltage (Capacity Voltage)
Battery temperature (Max Temp)
Discharge capacity (Discharge Ah)
Rest
Specified time after discharge end (Rest Time)
Rest completion condition of the battery temperature (Rest Temp)
Constant power-
constant voltage
discharge
CP - CV
Setting
Constant power value (Wattage)
Limit current (Limit Current)
Constant voltage value (CV Voltage)
Constant cell voltage value (Cell CV Voltage)
2
Cutoff
condition
Time since discharge start (Discharge Time)
Discharge capacity (Discharge Ah)
Battery temperature (Max Temp)
Specified time after constant voltage operation (CV Time)
Specified current after constant voltage operation (It Current)
Specified time after current falls below It Current (It Time)
Rest
Specified time after discharge end (Rest Time)
Rest completion condition of the battery temperature (Rest Temp)
Содержание PFX2500 Series
Страница 12: ...12 PFX2512 This page is intentionally blank ...
Страница 27: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2512 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2512 s specifications and outline drawing ...
Страница 96: ...96 PFX2512 This page is intentionally blank ...
Страница 140: ...140 PFX2512 Troubleshooting This page is intentionally blank ...
Страница 143: ......