4-8 Remote Control
PAS SERIES
4.1.3 Controlling the Output Voltage Using External
Resistance
This method is used to control the output voltage using an external
resistance (Rext) in the range 0 k
Ω
to approx. 10 k
Ω
.
• The insulation of the external resistor (Rext) and the
cable connecting it should be greater than the isola-
tion voltage of the unit. For details on the isolation
voltage of each model, see Chapter8 "Specifica-
tions".
• When using shielded wires for connection, protect
You can select the following two types of modes through CONFIG
settings. For details on CONFIG settings, see "3.2.4 Unit Configura-
tion (CONFIG)".
Press the SET switch to check the preset voltage.
10 k
Ω
→
MAX OUT CV mode
The output voltage (Eo) varies in the range of 0 to rated output
voltage by setting the external resistance (Rext) in the range of
0 k
Ω
to 10 k
Ω
.
Eo = Ertg•Rext /10 [V]
Rext = 10•Eo /Ertg [V]
Ertg: Rated output voltage [V]
10 k
Ω
→
0 OUT CV mode (FAIL SAFE)
The output voltage (Eo) varies in the range of rated output voltage
to 0 by setting the external resistance (Rext) in the range of 0 k
Ω
to 10 k
Ω
.
Eo =Ertg•(10-Rext)/10 [V]
Rext = 10•(Ertg-Eo)/Ertg [V]
Ertg: Rated output voltage [V]
WARNING
Содержание PAS 10-105
Страница 10: ...VIII PAS SERIES This page has been intentionally left blank ...
Страница 54: ...3 22 Basic Operation PAS SERIES This page has been intentionally left blank ...
Страница 78: ...4 24 Remote Control PAS SERIES This page has been intentionally left blank ...
Страница 131: ...PAS SERIES Specifications 8 13 This page has been intentionally left blank ...
Страница 137: ...PAS SERIES Specifications 8 19 This page has been intentionally left blank ...
Страница 145: ......