2-2 Before Using the Unit
PAS SERIES
Fig.2-1 Load current with peaks
Fig.2-2 Pulse-shaped load cur-
rent
2.2.2 When the Load Generates a Reverse Current to
the Power Supply
The unit cannot absorb a reverse current from a regenerative load
such as an inverter, converter, or transformer that supplies current to a
power supply. Consequently, the output voltage will increase and the
output will fluctuate.
For these types of loads, connect a resistor R
D
as shown in Fig.2-3 to
bypass the reverse current. However, the current capacity to the load
decreases by Irp.
•
For resistor R
D
, select an appropriate resistor rated
for the power (allowing sufficient margin).
If a resistor with insufficient rated power for the circuit
is used, it may burn out.
Fig.2-3 Remedy for regenerative load
Preset constant current value
Meter indication value (mean value)
Preset constant current value
Meter indication value (mean value)
CAUTION
I
O
R
D
E
O
Equivalent circuit of the unit
Regenerative load
Load
+
0
Reverse current
Output current
waveform
-I
O
+I
O
I
rp
R
D
[
ȍ
]
E
O
[V]
I
rp
[A]
R
D
: Reverse current bypass dummy load
E
O
: Output voltage
I
rp
: Max. reverse current
Содержание PAS 10-105
Страница 10: ...VIII PAS SERIES This page has been intentionally left blank ...
Страница 54: ...3 22 Basic Operation PAS SERIES This page has been intentionally left blank ...
Страница 78: ...4 24 Remote Control PAS SERIES This page has been intentionally left blank ...
Страница 131: ...PAS SERIES Specifications 8 13 This page has been intentionally left blank ...
Страница 137: ...PAS SERIES Specifications 8 19 This page has been intentionally left blank ...
Страница 145: ......