4-40
GPIB Operation
KJM6335
Details of Event Status Register and Event Status Enable
Register
Bit
Register name
Explanation
7
Not used with the KJM6335.
6
Not used with the KJM6335.
5
CME (Command Error)
Any one of the following events has been encountered while
decoding a message:
• Syntax error in the message received
• Illegal character data received
• Illegal suffix unit received
• Illegal data type received
4
EXE (Execution Error)
Any one of the following events has been encountered while
running a message:
• Received data out of bounds
• Received message not supported
3
Not used with the KJM6335.
2
Not used with the KJM6335.
1
UIS (Undesirable Input Signal)This register is set in case of any one of the following event
• No signal is input
• The clock frequency is outside the specified range
0
IMR (In Meas Range)
This register is set when the measured value decreases to 20
or less.
Table 4-22 Event Status Register and Event
Status Enable Register
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
NOTE
• The individual bits of the event status register and the
event status enable register are set when they are 1 and are
reset when they are 0.
• Run *ESR? to read the event status register and *CLS to
reset it.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Содержание KJM6335
Страница 1: ...KJM 6335 Part No Z1 002 622 IA002856 Jul 2009 KJM6335 OPERATION MANUAL CD TIME INTERVAL JITTER METER...
Страница 10: ...VIII Contents KJM6335...
Страница 12: ...P 2 Preface KJM6335...
Страница 41: ...KJM6335 Operation 3 7...
Страница 47: ...KJM6335 Operation 3 13...
Страница 104: ...4 46 GPIB Operation KJM6335...
Страница 112: ...5 8 Control Using EXT I O KJM6335...
Страница 126: ...A 2 Appendix KJM6335...