3-4
Operation
KJM6335
3.4
Operation
3.4.1 Measuring to Use RF Signal after Slicing and
Regenerated Clock Signal
CD player or other devices
KJM6335
RF signal after slicing
Clock signal
Clock-
regeneration
circuit
Slicer
Pick-up
RF
CLK
INHIBIT
Fig. 3-1
Connection
Connect an RF signal after slicing and clock signal to the
instrument from the CD player or other device, as shown in Fig. 3-
1. When carrying out measurement, note the signal output
impedance of the device under test. For example, if the signal
output impedance of the DUT is 50
Ω
, the input impedance of the
instrument should be set to 50
Ω
and it is recommended that a 50
Ω
coaxial cable (such as 3D-2V) be used. If you use a 10:1
oscilloscope probe, set the input impedance of the instrument to 1
M
Ω
.
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
NOTE
• When using the 10:1 probe, calibrate the probe to see 3.10,
"Calibrating the Probe".
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Содержание KJM6335
Страница 1: ...KJM 6335 Part No Z1 002 622 IA002856 Jul 2009 KJM6335 OPERATION MANUAL CD TIME INTERVAL JITTER METER...
Страница 10: ...VIII Contents KJM6335...
Страница 12: ...P 2 Preface KJM6335...
Страница 41: ...KJM6335 Operation 3 7...
Страница 47: ...KJM6335 Operation 3 13...
Страница 104: ...4 46 GPIB Operation KJM6335...
Страница 112: ...5 8 Control Using EXT I O KJM6335...
Страница 126: ...A 2 Appendix KJM6335...