KJM6335
Control Using EXT I/O
5-7
5.6
Judgement Output
The judgement of whether the measured value is larger or smaller
than the JUDGE level set on the panel will be output to the GO or
NOGO terminal of the EXT I/O connector.
Condition
G O
N O G O
Set value < measured value
L
H
Measured value > set value
H
L
Table 5-3
5.7
Within-Measuring-Range Output
An indication of whether the measured value is within the
measuring range of the instrument is output to the IN MEAS
RANGE terminal of the EXT I/O connector. The measuring range
of the instrument is within 20 %. If the measured value exceeds 20
%, a "Low" signal is output. If it is 20 % or less, a "High" signal is
output.
Содержание KJM6335