3-22
Operation
KJM6335
3.9
Selection of Trigger Edge
This meter measures the time difference between the RF signal and
the clock signal and indicates its dispersion as a standard deviation
value. To measure the time difference, specify the two edges of
each signal to be measured, using the RF SIGNAL TRIG EDGE
key and CLOCK TRG EDGE key.
Setting
RF signal
CLK signal
Setting
RF signal
CLK signal
RF
CLK
RF
CLK
Setting
RF signal
CLK signal
RF
CLK
Setting
RF signal
CLK signal
RF
CLK
Setting
RF signal
CLK signal
RF
CLK
Setting
RF signal
CLK signal
RF
CLK
Measure the time difference between the
rise of the RF signal and the rise of the
clock signal. Then, determine the
amount of jitter.
Measure the time difference between the
rise of the RF signal and the fall of the
clock signal. Then, determine the
amount of jitter.
Measure the time difference between the
rise of the RF signal and the rise of the
clock signal and that between the fall of
the RF signal and the rise of the clock
signal. Then, determine the amount of
jitter.
Measure the time difference between the
rise of the RF signal and the fall of the
clock signal and that between the fall of
the RF signal and the fall of the clock
signal. Then, determine the amount of
jitter.
Measure the time difference between the
fall of the RF signal and the rise of the
clock signal. Then, determine the
amount of jitter.
Measure the time difference between the
fall of the RF signal and the fall of the
clock signal. Then, determine the
amount of jitter.
Fig.3-17
Содержание KJM6335
Страница 1: ...KJM 6335 Part No Z1 002 622 IA002856 Jul 2009 KJM6335 OPERATION MANUAL CD TIME INTERVAL JITTER METER...
Страница 10: ...VIII Contents KJM6335...
Страница 12: ...P 2 Preface KJM6335...
Страница 41: ...KJM6335 Operation 3 7...
Страница 47: ...KJM6335 Operation 3 13...
Страница 104: ...4 46 GPIB Operation KJM6335...
Страница 112: ...5 8 Control Using EXT I O KJM6335...
Страница 126: ...A 2 Appendix KJM6335...