Specifications
7-1
Measurement Principle
Time difference in a clock signal is measured several times based
on an RF signal, and variations in the time difference are obtained
as percentages relative to a single clock period, assuming that a
single clock period is 100 %.
In the actual method of obtaining a standard deviation value, a time
difference is converted into a voltage and this voltage is converted
into the jitter value by the rms conversion circuit.
The unit of jitter value is % or ns.
RF signal
CLOCK signal
Time difference
Duration of one period
is regarded as 100 %.