16 Names and Functions of Parts
KDS6-0.2TR
[3]
Output connector
This is a voltage output and sensing terminal.
For this connection, use the D-Sub 25-pin male connector supplied.
To minimize the effects of ambient noise, read Chapter 3, "Basic Operations and
Connections" before using this connector.
[4]
Output ammeter
Shows the output current of the channel currently selected. The [----] display indi-
cates that the maximum permissible current has been exceeded. If the KDS6-0.2TR
is left in this state, the OCP function will activate after approximately 1.5 seconds.
For details, refer to "OCP function" on page 22.
[5]
Voltage setting display
Indicates the voltage setting of the channel currently selected.
[6]
RANGE key
Switches the current range of Channel 1. (Channels 2 and 3 are fixed at the 30-mA
range only)
[7]
CH SELECT key
Selects the channel for which voltage is to be set or output current is to be displayed.
[8]
Voltage control knob
Adjusts the voltage of the channel currently selected. The faster the knob is turned,
the greater the change in voltage value.
[9]
LOCAL key
Pressing this key when the RMT lamp remains lit (i.e., when the KDS6-0.2TR is
controlled through GPIB or RS-232C) enables the user to operate the KDS6-0.2TR
from the panel.
[10] SHIFT key
When the LOCAL key or CH SELECT key is pressed with this key held down, the
following key operations are executed:
SHIFT + LOCAL keys
Key lock is set or cleared.
SHIFT + CH SELECT keys
The sampling rate of the output ammeter is changed.
Содержание KDS6-0.2TR
Страница 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Страница 4: ...ii KDS6 0 2TR ...
Страница 8: ...4 Safety Precautions KDS6 0 2TR ...
Страница 12: ...8 Introduction KDS6 0 2TR ...
Страница 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Страница 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Страница 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Страница 58: ...54 Index KDS6 0 2TR ...
Страница 59: ......