24 Basic Operations and Connections
KDS6-0.2TR
Connection diagram
The sensing point should be located as close to the load end as possible.
If the sensing point is located far from the load end, hum may be picked up, or
imprecise voltage may be delivered on the load end, due to the voltage drop caused
by the resistance of the load cable.
Moreover, to minimize the effect of external noise, use twisted pair cables for the
load cables and sensing wires. It is also recommended that you twist the cabling of
each channel further before connecting it to the load.
Fig. 3-1
Connection to the load (example of good cabling)
Fig. 3-2
Connection to the load (example of poor cabling)
SLOW FAST
SLOW FAST
Load
Load
Load
COM/COM S
CH1+/CH1+ S
CH2+/CH2+ S
CH3+/CH3+ S
Load and COM
should be
connected at one
point.
Sensing point located close to the load end
Twist these four cables further.
SLOW FAST
SLOW FAST
Sensing point located far from the load
Load
Load
Load
The added
resistance of these
cables translates
into voltage error.
COM/COM S
CH1+/CH1+ S
CH2+/CH2+ S
CH3+/CH3+ S