KDS6-0.2TR
Remote Control 29
4.
Press the SHIFT key and save the newly established bit rate.
Pressing the SHIFT key saves the bit rate you have set. If you turn the power
off without pressing the SHIFT key, the setting will be lost.
RS-232C flow control
Xon/Xoff control is carried out using DC (Device Control) codes.
Fig. 4-1
Transmission control between the RS-232C terminal and
KDS6-0.2TR
RS-232C cross cable
Use a D-Sub 9-pin female-female cross cable (using #4-40UNC inch-system screws
as fixing screws) to link the computer and the KDS6-0.2TR.
Fig. 4-2
Pin assignment of RS-232C connector
Function
ASCII code
DC1
Transmission request
11 h
DC3
Transmission stop request
13 h
DC3
RXD
DC1
TXD
After receiving DC3, the KDS6-0.2TR pauses transmission within 3 characters.
Within 3 characters
Pause
Resume transmission
Resume transmission
Transmission control from the RS-232C terminal to the KDS6-0.2TR
DC3
TXD
RXD
The RS-232C terminal must pause transmission within
10 characters after receiving DC3.
Transmission control from the KDS6-0.2TR to the RS-232C terminal
DC1
Pause
Within 10 characters
#4-40UNC
Inch screw
#4-40UNC
Inch screw
1: CD (carrier detect)
2: RXD (receive data)
3: TXD (transmit data)
4: DTR (data terminal ready)
5: GND (signal ground)
6: DSR (data set ready)
7: RTS (request to send)
8: CTS (clear to send)
9: RI (ring indicator)
As viewed from the rear panel of the KDS6-0.2TR
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
D-sub 9-pin female
D-sub 9-pin female
Cross cable example
Содержание KDS6-0.2TR
Страница 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Страница 4: ...ii KDS6 0 2TR ...
Страница 8: ...4 Safety Precautions KDS6 0 2TR ...
Страница 12: ...8 Introduction KDS6 0 2TR ...
Страница 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Страница 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Страница 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Страница 58: ...54 Index KDS6 0 2TR ...
Страница 59: ......