32 Remote Control
KDS6-0.2TR
SILent (RS-232C only)
This specifies whether an acknowledge message is to be returned to a message
accompanied by a response message terminator.
This device message function is effective for the RS-232C only.
Either “OK” or “ERROR” is returned as the acknowledge message.
Program message
•
Syntax
Command message:
SILent <{ON|OFF|1|0}>
SIL <{ON|OFF|1|0}>
•
Program data
(Example) To set the device to return an acknowledge message
SIL 0
*RST
The following items are initialized by the *RST command message. For setting
details, refer to Table 4-2.
Program message
•
Syntax
Command message:
*RST
Table 4-2
System setting for executing the *RST command
GPIB addresses, RS-232C bit rates, and settings performed via SILent and *SRE
device messages are not initialized.
Setting acknowledge message
Data format
Character
1: ON (No message returned)
0: OFF (Message returned)
Initial value
1: ON (No message returned)
* The setting is reset to 1 when the power is turned
off.
Item
Setting value
Voltage setting (all channels)
0.0000 V
Channel selection
Channel 1
Channel-1 range
200 mA
Output ammeter sampling rate
Normal mode
OUTPUT
Off
KEYLOCK
Clear
Содержание KDS6-0.2TR
Страница 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Страница 4: ...ii KDS6 0 2TR ...
Страница 8: ...4 Safety Precautions KDS6 0 2TR ...
Страница 12: ...8 Introduction KDS6 0 2TR ...
Страница 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Страница 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Страница 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Страница 58: ...54 Index KDS6 0 2TR ...
Страница 59: ......