KDS6-0.2TR
Basic Operations and Connections 23
3.3
Connection to the Load
Pin assignment of output connector
The KDS6-0.2TR employs a D-Sub 25-pin female connector for the output connec-
tor.
To connect the KDS6-0.2TR to its load, connect to the output with the D-Sub 25-pin
male connector supplied, using Table 3-2 as a guide.
Table 3-2
Output connector pin assignment
Connector fixing screws
The fixing screws used for the output connector are metric screws (size M2.6).
• Operate the KDS6-0.2TR only with the sensing wires connected. Failure to con-
nect the sensing wires will impair the KDS6-0.2TR’s ability to deliver precise
voltages.
No.
Name
Remarks
No.
Name
Remarks
1
CH1 +
Channel-1 voltage output terminal
14
COM
Common output terminal
2
CH1 +S
Channel-1 sensing terminal
15
COM S
Common sensing terminal
3
NC
Not in use
16
NC
Not in use
4
CH2 +
Channel-2 voltage output terminal
17
NC
Not in use
5
CH2 +S
Channel-2 sensing terminal
18
NC
Not in use
6
NC
Not in use
19
NC
Not in use
7
CH3 +
Channel-3 voltage output terminal
20
NC
Not in use
8
CH3 +S
Channel-3 sensing terminal
21
NC
Not in use
9
NC
Not in use
22
NC
Not in use
10
NC
Not in use
23
NC
Not in use
11
NC
Not in use
24
NC
Not in use
12
NC
Not in use
25
NC
Not in use
13
NC
Not in use
As viewed from the front panel of the KDS6-0.2TR
1
2
3
4
5
6
7
20 19 18 17 16 15 14
25 24 23 22 21
13 12 11 10
9
8
CH3 +
NC
CH2 +S
CH2 +
NC
CH1 +S
CH1 +
COM
COM S
NC
NC
NC
NC
CH3 +S
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
NC
M2.6
Metric screw
M2.6
Metric screw
NOTE
Содержание KDS6-0.2TR
Страница 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Страница 4: ...ii KDS6 0 2TR ...
Страница 8: ...4 Safety Precautions KDS6 0 2TR ...
Страница 12: ...8 Introduction KDS6 0 2TR ...
Страница 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Страница 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Страница 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Страница 58: ...54 Index KDS6 0 2TR ...
Страница 59: ......