Disassembly and Repair
5
Keysight U3401A User’s and Service Guide
87
To replace the power line fuse
The power line fuse is located within the multimeter’s fuse-holder assembly on
the rear panel. The multimeter is shipped from the factory with a power-line fuse
installed (according to country of destination). See
. If you determine
that the fuse is faulty, replace it with one of the same size and rating.
Table 5-1
Type of supplied fuse (according to country of destination)
Type of fuse (time-lag, low breaking fuse)
Input line voltage
Part number
0.25A, 250V, 5x20 mm
100 V to 120 V
A02-62-25592-3U
0.125A, 250V, 5x20 mm
220 V to 240 V
A02-62-25648-1U
2
Remove the line voltage selector from
fuse holder assembly.
1
Disconnect the power cord. Depress tab 1 and
2 and pull the fuse holder from rear panel.
1
2
4
Replace fuse holder assembly in
rear panel.
3
Rotate the line voltage selector and
reinstall it back, so that the correct voltage
appears in the fuse holder window.