Operations and Features
2
Keysight U3401A User’s and Service Guide
59
Using the setup menu
The Setup menu allows you to customize a number of non-volatile instrument
configurations. The content of the Setup menu are shown in
.
Table 2-3
Setup menu and communications parameters
First tier
menu
Description
Second tier
menu
Description
Default
factory
setting
Available settings
rS232
RS232 interface
parameters
bAUd
Baud rate for remote
communication with a PC (remote
control).
9600
9600, 4800, 2400, 1200,
600 and 300
PArtY
Parity bit for remote
communication with a PC.
None
None, odd or even
dAtA
Data bit length for remote
communication with a PC.
8
7 or 8
StoP
Stop bit.
1
1 or 2
Echo
ECHO. Return a character to PC in
remote communication.
OFF
ON or OFF
Printer-Only. Print measured data
to a PC in remote communication.
OFF
ON or OFF
Ref
Ω
Reference
impedance
600
Ω
8000
Ω
, 1200
Ω
, 1000
Ω
,
900
Ω
, 800
Ω
, 600
Ω
,
500
Ω
, 300
Ω
, 250
Ω
,
150
Ω
, 135
Ω
, 125
Ω
,
124
Ω
, 110
Ω
, 93
Ω
, 75
Ω
,
50
Ω
, 16
Ω
, 8
Ω
, 4
Ω
, 2
Ω
bEEP
Beeper selection
bEEP
Enable or disable beep operation.
ON
ON or OFF
bEEPF
Beep frequency.
4096 Hz
4096, 2048, 1024 8192 Hz
rHold
Refresh hold
StAtE
State.
OFF
ON or OFF
Count
Variation count.
300
100, 200, 300, 400, 500,
600, 700, 800, 900, 1000.
Notes:
[a] For calibration use only.
[b] Variation count is used to recognize new value will be updated once the variation of measuring value is esceeded the setting value.
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...