Test Considerations
For optimum performance, all procedures should comply with the following
recommendations:
– Ensure that the calibration ambient temperature is stable and between 18°C
and 28°C. Ideally the calibration should be performed at 23°C ±1°C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 30 minutes warm-up period with a Shorting Plug connected to the Hi
and Lo input terminals.
– Use shielded twisted pair PTFE-insulated cables to reduce settling and noise
errors. Keep the input cables as short as possible.
– Connect the input cable shields to earth ground. Connect the calibrator Lo
source to earth ground at the calibrator except where noted in the procedures.
It is important that the Lo to earth ground connection be made at only one
place in the circuit to avoid ground loops.
NOTE
Please ensure that the calibration standards and test procedures used do not
introduce additional errors. Ideally, the standards used to verify and adjust the
instrument should be an order of magnitude more accurate than each
instrument range full-scale error specification.
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...