1
Getting Started
36
Keysight U3401A User’s and Service Guide
DC Voltage
– Five ranges: 500.00 mV, 5.00 V, 50.00 V, 500.00 V, 1000.00 V
– Measurement method: Sigma Delta A-to-D converter
– Input impedance: 10 M
Ω
±2% range (typical)
– Input protection: 1000 V on all ranges
1
Press .
2
Connect the red and black test leads to the respective input terminals as
shown in
3
Probe the test points and read the display. In autoranging mode, the
multimeter automatically selects the appropriate range and the measurement
is displayed.
Figure 1-11
DCV terminal connection and display
+
DCV
voltage
source
–