1
Getting Started
20
Keysight U3401A User’s and Service Guide
Connecting power to the instrument
Connect the power cord and press the power switch to turn on the multimeter.
The front panel display illuminates while the multimeter performs its power-on
self-test. (If the multimeter does not power-on, refer
page 84). During the power-on session, press
to hold the full display. Press
any key to resume the power-on self-test.
The multimeter powers up in the DC voltage function with autoranging enabled.
If self-test is successful, the multimeter goes to normal operation.
If the self-test
fails, either a full annunciator or a blank display is displayed without entering the
normal operation. If the self-test repeatedly fails, contact your nearest Keysight
Sales and Service Office.
NOTE
The multimeter will operate at any line voltage between 90 VAC and 264 VAC
when the line voltage selector is set properly with frequency range 50 Hz or 60 Hz.
CAUTION
– Before turning on the multimeter, make sure the line voltage selector is
set to the correct position for the applied line voltage to the power line
cord connector.
– Do not apply a line voltage that exceeds the specified range of the line
cord connector.
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...