Operations and Features
2
Keysight U3401A User’s and Service Guide
47
dBm
The logarithmic dBm (decibels relative to one milliwatt) scale is often used in RF
signal measurements. The multimeter’s dBm operation takes a measurement and
calculates the power delivered to a reference impedance (typically 50, 75 or 600
Ω
)
.
The formula used for conversion from the voltage reading is:
dBm = 10 x Log
10
[1000 x (Reading
2
/ reference impedance)]
Figure 2-1
Typical dBm operation display
The default reference impedance value is 600
Ω
.. You can select the following
21 reference impedance values:
2
Ω
, 4
Ω
, 8
Ω,
16
Ω
, 50
Ω
, 75
Ω
, 93
Ω
, 110
Ω
, 124
Ω,
125
Ω
, 135
Ω
, 150
Ω
, 250
Ω,
300
Ω,
500
Ω,
600
Ω
, 800
Ω
, 900
Ω
, 1000
Ω
, 1200
Ω
, 8000
Ω
.
The dBm operation can be applied to DCV, ACV and DCV + ACV measurement
functions only. The multimeter displays the dBm modifier on the primary display
and displays the reference impedance selection on the secondary display.
Procedure 1
1
Press
to enable the dBm modifier mode and dBm operation.
2
Use
and
to select the desired reference impedance.
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...