Operations and Features
2
Keysight U3401A User’s and Service Guide
61
Trigger mode
This multimeter has two types of trigger mode.
– Internal — triggers measurements continuously.
– External — triggers a measurement only when a direction is given.
The external trigger is used with a delay settling that has been set by the
multimeter automatically. The amount of trigger delay varies depending on
different function. When external trigger is enabled, the multimeter determines
the range for the primary display based on the input at that time.
Autoranging response time is required before a reading is displayed when you
change the input after a trigger is received.
Procedure
1
Press
to enter trigger mode and standby condition. The
TRIG
annunciator will be lit and display is indicated with “–––––”.
2
Press
to get a new value. After a measurement, the result will be
indicated and held on the display.
3
Press
to get a new value again.
4
Press
,
or
to select autoranging or manual range as
necessary.
5
Press
to disable external trigger operation mode.
Figure 2-10
Typical trigger mode display
Содержание U3401A
Страница 1: ...Keysight U3401A 4 1 2 Digit Dual Display Multimeter User s and Service Guide ...
Страница 12: ...12 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...3 Application Tutorial 70 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 82: ...4 Performance Test 82 Keysight U3401A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...