1
Getting Started
32
Keysight U3401A User’s and Service Guide
Table 1-3
Input terminal for different functions
Measurement function
Input terminal
Overload protection
DC voltage (VDC)
COM
1000 VDC
AC voltage (VAC), frequency (Hz)
750 VAC RMS, 1100 V PEAK,
2x10
7
V-Hz normal mode, or
1x10
6
V-Hz common mode
Miliampere (mA), frequency (Hz)
mA
500 mADC or AC RMS
10 A, frequency (Hz)
10 A
10 ADC or AC RMS continuous, and
>10 A DC RMS for 20 seconds maximum
Resistance (
Ω
)
500 VDC or AC RMS
Diode test, continuity test
500 VDC or AC RMS
All functions
Any terminal to earth
1000 VDC or AC PEAK