EO measurements (lasers, modulators)
The measurement of an EO transducer is a combination of
input modulating current and output optical modulation
power. Slope responsivity is used to describe how a change
in input current produces a change in optical power.
Graphically this is shown in the figure below.
P
out
mW
I
in
mA
Responsivity R
s
W A
P
out
I
in
----------------
=
R
s
dB
20log
10
R
s
W A
W A
----------------------------
=
An LCA measures input modulating current and output
modulation power and displays the ratio of the two in
Watts/Amp, either linearly or in decibels.
Содержание N4375D
Страница 1: ...Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ...1 ...
Страница 9: ...Keysight N4375D Lightwave Component Analyzer User s Guide 7 Ordering Information 165 Index 169 ...
Страница 10: ......
Страница 34: ......
Страница 44: ......
Страница 78: ......
Страница 124: ......
Страница 131: ...Definition of Terms 10 Definition of Terms Definition of Terms 130 ...
Страница 140: ......
Страница 146: ...Specifications 144 Keysight N4375D Lightwave Component Analyzer User s Guide Transmitter and Receiver Specifications ...
Страница 154: ...Specifications 152 Keysight N4375D Lightwave Component Analyzer User s Guide General Characteristics ...
Страница 155: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 153 ...
Страница 157: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 155 For 2 port PNA Options 22x ...
Страница 158: ...Specifications 156 Keysight N4375D Lightwave Component Analyzer User s Guide For 4 port PNA Options 42x ...
Страница 166: ......
Страница 167: ...Ordering Information 13 Ordering Information ...
Страница 168: ...Ordering Information 166 Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 170: ......
Страница 173: ......