Performance Quick Check
40
Keysight
N4375D Lightwave Component Analyzer, User’s Guide
6. Select the OO measurement tab.
7. Set the following parameters:
• If you have option 102 (Dual wavelength 1310 nm,
1550 nm), set Wavelength to the value at which you are
testing your DUTs (1310 nm or 1550 nm).
• Laser power: +4 dBm
• Optical input: input1 (“+7 dBm max”) or input 2
(“+17 dBm max”)
• User Calibration: None
• Modulator optimization: Continuous
• Forward RF Power: +5 dBm
8. Click
, then
.
9. Right click the measurement trace, and select “Autoscale
All” in the context menu.
At the end of the measurement, there should be a flat trace.
• An ideal trace might
look like this.
Содержание N4375D
Страница 1: ...Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ...1 ...
Страница 9: ...Keysight N4375D Lightwave Component Analyzer User s Guide 7 Ordering Information 165 Index 169 ...
Страница 10: ......
Страница 34: ......
Страница 44: ......
Страница 78: ......
Страница 124: ......
Страница 131: ...Definition of Terms 10 Definition of Terms Definition of Terms 130 ...
Страница 140: ......
Страница 146: ...Specifications 144 Keysight N4375D Lightwave Component Analyzer User s Guide Transmitter and Receiver Specifications ...
Страница 154: ...Specifications 152 Keysight N4375D Lightwave Component Analyzer User s Guide General Characteristics ...
Страница 155: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 153 ...
Страница 157: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 155 For 2 port PNA Options 22x ...
Страница 158: ...Specifications 156 Keysight N4375D Lightwave Component Analyzer User s Guide For 4 port PNA Options 42x ...
Страница 166: ......
Страница 167: ...Ordering Information 13 Ordering Information ...
Страница 168: ...Ordering Information 166 Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 170: ......
Страница 173: ......