Making Measurements
102
Keysight
N4375D Lightwave Component Analyzer, User’s Guide
• If for “User Calibration”, you select “None”, the LCA
uses the factory default calibration.
Optical path and electrical path deembedding is not applied during a user
calibration.
• The factory calibration ensures the uncertainty
specified, but particularly for improved results at lower
frequencies, we recommend an extra calibration. In this
case, for “User Calibration”, select “Measure”.
• If you want to reuse and existing, saved calibration, for
“User Calibration”, select “Load”, and select the file
from the Windows Explorer window.
9 Click
to start the measurement preparation.
If you are calibrating, when the following screen is
displayed, make sure the optical output is connected to
the optical input using the patchcord supplied.
The patchcord has the following part number:
• Single- mode angled/angled connector: N4373- 87906
• Single- mode angled/straight connector: N4373- 87907
Please also refer to the documentation supplied with the
patchcord.
N O T E
Содержание N4375D
Страница 1: ...Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ...1 ...
Страница 9: ...Keysight N4375D Lightwave Component Analyzer User s Guide 7 Ordering Information 165 Index 169 ...
Страница 10: ......
Страница 34: ......
Страница 44: ......
Страница 78: ......
Страница 124: ......
Страница 131: ...Definition of Terms 10 Definition of Terms Definition of Terms 130 ...
Страница 140: ......
Страница 146: ...Specifications 144 Keysight N4375D Lightwave Component Analyzer User s Guide Transmitter and Receiver Specifications ...
Страница 154: ...Specifications 152 Keysight N4375D Lightwave Component Analyzer User s Guide General Characteristics ...
Страница 155: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 153 ...
Страница 157: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 155 For 2 port PNA Options 22x ...
Страница 158: ...Specifications 156 Keysight N4375D Lightwave Component Analyzer User s Guide For 4 port PNA Options 42x ...
Страница 166: ......
Страница 167: ...Ordering Information 13 Ordering Information ...
Страница 168: ...Ordering Information 166 Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 170: ......
Страница 173: ......