Making Measurements
Keysight
N4375D Lightwave Component Analyzer, User’s Guide
83
1 Set your measurement parameters.
2 From the “Calibration” menu, select “Calibration Wizard”.
3 Select “Use Electronic Calibration”, and click [Next>].
4 Make sure “Show Advanced Settings” is selected.
5 Select “3 Port Ecal”, Select 3 Port ECal Configuration
according to the screen shot.
6 Click the [Next>] button.
7 Select “Thru #1 and “Thru #2” according to the screen
shot.
8 Select the appropriate Cal Kit for each port, or “ECal
Thru as Unknown”.
9 Make sure “Do Orientation” is selected, and click [Next >].
10 When using the 2 port ECal module connect the ECal
Module to Port A and the RF Cables to NWA ports 3 and
2, respectively.
Содержание N4375D
Страница 1: ...Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ...1 ...
Страница 9: ...Keysight N4375D Lightwave Component Analyzer User s Guide 7 Ordering Information 165 Index 169 ...
Страница 10: ......
Страница 34: ......
Страница 44: ......
Страница 78: ......
Страница 124: ......
Страница 131: ...Definition of Terms 10 Definition of Terms Definition of Terms 130 ...
Страница 140: ......
Страница 146: ...Specifications 144 Keysight N4375D Lightwave Component Analyzer User s Guide Transmitter and Receiver Specifications ...
Страница 154: ...Specifications 152 Keysight N4375D Lightwave Component Analyzer User s Guide General Characteristics ...
Страница 155: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 153 ...
Страница 157: ...Specifications Keysight N4375D Lightwave Component Analyzer User s Guide 155 For 2 port PNA Options 22x ...
Страница 158: ...Specifications 156 Keysight N4375D Lightwave Component Analyzer User s Guide For 4 port PNA Options 42x ...
Страница 166: ......
Страница 167: ...Ordering Information 13 Ordering Information ...
Страница 168: ...Ordering Information 166 Keysight N4375D Lightwave Component Analyzer User s Guide ...
Страница 170: ......
Страница 173: ......