Step 5: Verify Operation of the Agilent M9392A PXI Vector Signal Generator
Conduct M9392A Field Calibration
Required Hardware
Please refer to the following table for recommended hardware. Additionally, some high-quality flexible 3.5 mm
cables will be required, plus adapters to connect the PSG (precision signal generator) and power meter to 3.5 mm
connectors.
Hardware
Description
Agilent E8257D-532 or E8267D-532
Precision Signal Generator 31.8 GHz
Agilent N1913A or N1914A
Single / Dual Channel Power Meter
Agilent N8485A, Option 100
10 MHz to 26.5 GHz Power Sensor
Agilent 11730A
Power Sensor Cable
A Field Calibration should be performed upon initial setup and whenever one of more of the modules that compose
the M9392A PXI Vector Signal Analyzer is removed, reconfigured into different slots, or when cables are dis-
connected or changed. The Field Calibration should be repeated on a regular three month schedule, in order to
achieve the highest level of performance.
Perform an Agilent M9392A PXI Vector Signal Analyzer Field Calibration by using the M9392A SFP and the
included Help system.
1. Open the M9392A soft front panel (SFP).
2. Conduct a Field Calibration (
Utility > Field Calibration...
).
Use the
Help
button on the Agilent M9392A Field Calibration screen to access context-sensitive help to guide you
through the procedure.
Agilent M9392A PXI Vector Signal Analyzer Startup Guide
21
Содержание M9392A
Страница 1: ...Startup Guide Keysight M9392A PXI Vector Signal Analyzer 50 MHz to 26 5 GHz...
Страница 2: ......
Страница 7: ......
Страница 8: ...Documentation Map Documentation Map Agilent M9392A PXI Vector Signal Analyzer Startup Guide 7...
Страница 40: ...Streaming Data Overview System diagram Agilent M9392A PXI Vector Signal Analyzer Startup Guide 39...
Страница 48: ......