Streaming Data Overview
c. Close the EEPROM Programmer application, and all other applications.
d. Cycle power to the chassis.
If you are using a remote controller, Shut Down the PC
BEFORE
you power down the
chassis. When you restore power, power up the chassis
BEFORE
you power up the
PC.
e. Run the EEPROM Programmer application again, and load the license (*.epr) file sent to you by Agi-
lent Support (see Step 2 above). The license file update should only take several seconds.
5. To verify the streaming upgrade is complete and operating correctly, use the M9392A Soft Front Panel
(SFP) to conduct a streaming data capture test:
a. Open the M9392 SFP by selecting:
Start > All Programs > Agilent > M9392 > M9392 SFP
.
b. In the SFP, select the Streaming Enabled checkbox (
) to enable streaming.