Streaming Data Overview
c. Create a test.bin file to store the streaming data. To do this, click the
(File Save As) button.
In the Save As dialog, type in the file name “test.bin” and click Save.
d. Under Acquisition Mode, select Duration Limited.
e. Set the Acquisition Duration to 4 seconds.
f. If you ordered the M9202AU-V10 upgrade, set the bandwidth between 50 MHz and 100 MHz.
Otherwise, leave the bandwidth at the default of 40 MHz.
g. Click the Start button to initiate the data capture. When the streaming session is complete, the
status bar shows "Streaming Complete" and the Start button is re-enabled (Stop and Abort but-
tons are disabled).
You have completed the Streaming upgrade for your M9392A Vector Signal Analyzer and M9202A IF Dig-
itizer.
Streaming System Configuration
The Agilent M9392A streaming solution uses a controller disk drive, and (optionally) an external RAID solution.
Controller and RAID selection depends on the data capacity, and data read/write rate requirements. The following
are Agilent-recommended storage packages:
Agilent's predefined external RAID packages have been tested to guarantee sustained data rates
with the M9392A. These predefined mass storage packages ensure performance when used with the
recommended controllers.
l
Recommended controllers:
Controller
Capture Time
Disk Write Rate
Agilent
M9036A PXIe Embedded Controller
< 6 seconds @ 100 MHz BW
(depends on memory)
~ 40 MB/s
(assumes 4 GB RAM)
Desktop computer from the computer list at
http://cp-
.literature.agilent.com/litweb/pdf/5990-
7632EN.pdf
(for example, the Dell T3500 or
Dell T5500 computer)
~ 18 seconds @ 100 MHz BW
(depends on memory)
~ 40 to 100 MB/s
(depends on disk drive)
l
Agilent recommends the following
RAID storage solutions:
JMR Model
Capture Time
Disk Write Rate
AGIL-G4-8T
(HDD)
> 3 Hours @ 100 MHz BW
> 500 MB/s
AGIL-G4-16T
(HDD)
> 6 Hours @ 100 MHz BW
AGIL-G4-32T
(HDD)
> 13 Hours @ 100 MHz BW
AGIL-G4-0.4T
(SSD)
10 minutes @ 100 MHz BW
For details on connecting and configuring the above JMR RAID solutions to work with the M9392A,
refer to the white paper (located on the web at
http://cp.literature.agilent.com/litweb/pdf/5990-
). If you installed the M9392A software, you can access the white paper from
Start > All Programs > Agilent > M9392A > M9392A RAID White Paper
.
38
Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Содержание M9392A
Страница 1: ...Startup Guide Keysight M9392A PXI Vector Signal Analyzer 50 MHz to 26 5 GHz...
Страница 2: ......
Страница 7: ......
Страница 8: ...Documentation Map Documentation Map Agilent M9392A PXI Vector Signal Analyzer Startup Guide 7...
Страница 40: ...Streaming Data Overview System diagram Agilent M9392A PXI Vector Signal Analyzer Startup Guide 39...
Страница 48: ......