M9392A Troubleshooting
Status LED States
Once you open the M9392A SFP, select a module, and then choose
Connect
. The module is initial-
ized. Until it is initialized, the LED behavior is not entirely predictable.The front panel LED behav-
ior is valid only when the soft front panel (SFP) is running or when the
Initialize
function/method has been called, using the application programming interface (API).
Module
Green
Amber
Red
Off
M9202A
Indicates the trigger is
armed and waiting for a
valid trigger to occur.
Identifies the module asso-
ciated with the current
channel in the Soft Front
Panel when the “Finder”
button is depressed.
Indicates that the trigger
has occurred, the acqui-
sition is complete, and the
data is waiting to be read.
Indicates a failure in the
power supplies. Mod-
ule hardware health
can’t be determined
until the power supply
failure is resolved.
M9302A
All power supplies are
functioning within their tol-
erances, and all hard-
ware is functioning and a
10 MHz Ref In is con-
nected.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hard-
ware failures.
l
Indicates a failure in
the power supplies and
module hardware.
l
Indicates the absence
of the 10 MHz REF IN
signal or 10 MHz is
present, but SFP has
not initialized.
Indicates a failure in the
power supplies. Mod-
ule hardware health
can’t be determined
until the power supply
failure is resolved.
M9351A
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is func-
tioning.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hard-
ware failures.
l
Indicates a failure in
the power supplies and
module hardware.
l
*
When Meas
Frequency is ≥ 2.75
GHz
l
When the M9351A LO
IN is disconnected.
l
When the M9360A LO
IN is disconnected.
l
When the 100 MHz
REF IN is dis-
connected.
Indicates a failure in the
power supplies. Mod-
ule hardware health
can’t be determined
until the power supply
failure is resolved.
M9360A
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is func-
tioning.
Indicates that all power
supplies are functioning
within their tolerances, but
there may be other hard-
ware failures.
Indicates a failure in the
power supplies and module
hardware.
Indicates a failure in the
power supplies. Mod-
ule hardware health
can’t be determined
until the power supply
failure is resolved.
M9361A
Indicates that all power
supplies are functioning
within their tolerances,
and all hardware is func-
tioning.
Indicates that all power
supplies are functioning
within their tolerances, but
there are other hardware
failures.
Indicates a failure in the
power supplies and module
hardware.
Indicates a failure in the
power supplies, but
module hardware
seems to be func-
tioning. Module hard-
ware functioning
indication can be erro-
neous in this mode
since this circuitry can
be compromised by a
power supply failure.
*
Once the SFP is used, the LED may be red. When first plugged in, it may or may not be red.
24
Agilent M9392A PXI Vector Signal Analyzer Startup Guide
Содержание M9392A
Страница 1: ...Startup Guide Keysight M9392A PXI Vector Signal Analyzer 50 MHz to 26 5 GHz...
Страница 2: ......
Страница 7: ......
Страница 8: ...Documentation Map Documentation Map Agilent M9392A PXI Vector Signal Analyzer Startup Guide 7...
Страница 40: ...Streaming Data Overview System diagram Agilent M9392A PXI Vector Signal Analyzer Startup Guide 39...
Страница 48: ......