Using InfiniiMax III+ Probe Heads
2
InfiniiMax III+ Series Probes User’s Guide
73
Soldering Tips
When using the N5441A and N2836A solder-in heads and N5439A ZIF probe
heads, be sure to observe the following guidelines. For more information, refer to
the user’s guide.
• Use a temperature-controlled soldering iron station, if possible.
• Set the temperature of the soldering iron’s tip to between 370° C and 420° C
(for non RoHS standards).
• Use the smallest tip possible.
• Use an optical aid of some sort (microscope preferred).
• Employ minimal dwell times on the solder joint (< 2 seconds).
• Solder only the tip of the wire onto your DUT. The solder should not get close to
the existing solder ball on the tip.
• Use enough flux when soldering the tips into a DUT.
Figure 60
Applying Flux to Leads and Wires
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...