110
InfiniiMax III+ Series Probes User’s Guide
6
Performance Plots
Figure 86
Fidelity of Step Response (Probe System Versus Input Signal)
This example was for a 13 GHz system (N2832A probe) measuring a 6.6 GHz
signal. This ratio can be used to determine the measurement bandwidth needed
for other bandwidth signals (for a 3% rise time measurement error):
This equation can be used to determine the probe-scope system BW needed to
measure a certain BW signal.
BW
probe-scope system
13 GHz BW
signal
×
6.6 GHz
-------------------------------------------------
≅
BW
probe-scope system
1.97 BW
signal
×
≅
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...