98
InfiniiMax III+ Series Probes User’s Guide
5
Specifications and Characteristics
Input Referred Noise Spectral Density
Diff 5:1 attenuation 33.5 nV/rt(Hz),
Diff 10:1 attenuation 53.9 nV/rt(Hz),
SE A or B 5:1 attenuation 27.8 nV/rt(Hz),
SE A or B 10:1 attenuation 47.7 nV/rt(Hz),
CM 5:1 attenuation 21.8 nV/rt(Hz),
CM 10:1 attenuation 38.4 nV/rt(Hz)
Input Referred Noise Example
4.5 mV
rms
in diff mode 5:1 attenuation
with
¡
18 GHz probe head and N7001A
13 GHz probe amplifier
4.5 mV
rms
in diff mode 5:1
attenuation with 28 GHz N5444A
probe head and N7001A 13 GHz
probe amplifier
Maximum Input Voltage
18 V
peak
CAT 1
8 V
peak
without violating maximum
input power
Table 21
Probe Amplifier Characteristics (Sheet 2 of 2)
Item
N5439A, N5441A, N2836A, N2848A
With N5444A Probe Head
(2.92 mm, SMA, 3.5 mm)
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...