62
InfiniiMax III+ Series Probes User’s Guide
2
Using InfiniiMax III+ Probe Heads
Figure 47
Probe Head Connection to DUT (
Non-InfiniiMode
)
Ensuring maximum performance
The specifications and performance plots of the N2836A probe head were
measured with a nominal spacing of 40 mil (1 mm). In order to achieve the proper
response as shown in the performance plots, keep the mini-axial lead
resistors
roughly parallel as shown in
, and use the tip wires on the mini-axial
leads to get the desired span.
Figure 48
Proper Position of Resistors
If you need to position the mini-axial lead resistors different than shown in this
figure (that is, resistor bodies close together or spread way apart), use N2807A
and N2808A PrecisionProbe products to perform an AC calibration of the probe.
Содержание InfiniiMax III+ Series
Страница 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Страница 11: ...Using InfiniiMax III Series Probes 1 InfiniiMax III Series Probes User s Guide 11 Figure 5 N2830 1 2A Probes...
Страница 12: ...12 InfiniiMax III Series Probes User s Guide 1 Using InfiniiMax III Series Probes Figure 6 N7000 1 2 3A Probes...
Страница 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Страница 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Страница 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Страница 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Страница 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Страница 136: ...136 InfiniiMax III Series Probes User s Guide 7 Performance Verification for N2830A Series Probes...
Страница 150: ...150 InfiniiMax III Series Probes User s Guide 8 Performance Verification for N7000 Series Probes...
Страница 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...